GONIOMETRY
FOR WRIST & FINGERS
WHY HAND IS SO SPECIAL…
• 27 bones
• 27 Joints
• 34 muscles
• 3 major nerves
• 100 ligaments & tendons
• 2 major arteries
• Major representation in Brain
• & so on…..
THE REAL MAN…
HAND PHYSIOTHERAPY
– A SUPER SPECIALITY
• Micro & Macro surgeries – Vascular, Orthopaedic, Plastic surgeons, Neuro
surgeons etc
• Hand surgeons, Thumb surgeons etc.,
• Toe to thumb transfers
• Tendon repair & transplants
• Hand transplant – Transplants from cadaver.
• & Many more
HAND PHYSIOTHERAPY
UNIVERSAL FULL CIRCLE
GONIOMETER
UNIVERSAL FINGER GONIOMETER
ROLYAN FINGER/TOE GONIOMETER
DEVORE POCKET GONIOMETER
BASELINE DIGIT FINGER GONIOMETER
EXTENDABLE GONIOMETERS
NORMAL WRIST ROM
NORMAL FINGER ROM
NORMAL THUMB ROM
WRIST FLEXION
Test Position
• Subject sitting
• Forearm stabilized on table
• Flex wrist (fingers relaxed)
Goniometer Alignment
• Axis – lateral wrist
• Stationary arm – aligned with ulna
• Moving arm – aligned with fifth
metacarpal
Normal Range
• 75° ± 6.6° (American Academy of
Orthopaedic Surgeons)
• 60° (American Medical
Association)
• 76.4° (mean) 6.3° (standard
deviation), (Boone and Azen)
WRIST EXTENSION
Test Position
• Subject sitting
• Forearm stabilized on table
• Extended wrist (fingers relaxed)
Goniometer Alignment
• Axis – lateral wrist (triquetrum)
• Stationary arm – aligned with ulna
• Moving arm – aligned with fifth
metacarpal
Normal Range
74° ± 6.6° (American Academy of
Orthopaedic Surgeons)
60° (American Medical Association)
74.9° (mean) 6.4° (standard deviation),
(Boone and Azen)
WRIST RADIAL DEVIATION
Test Position
• Subject sitting with forearm resting on
table
• Stabilize forearm to prevent pronation
or supination
Goniometer Alignment
• Axis – capitate
• Stationary arm – aligned with forearm
(lateral epicondyle)
• Moving arm – aligned with metacarpal
of middle finger
Normal Range
• 21° ± 4.0° (American Academy of
Orthopaedic Surgeons)
• 20° (American Medical Association)
• 21.5° (mean) 4.0° (standard
deviation), (Boone and Azen)
WRIST ULNAR DEVIATION
Test Position
• Subject sitting with forearm resting on
table
• Stabilize forearm to prevent pronation
or supination
Goniometer Alignment
• Axis – capitate
• Stationary arm – aligned with forearm
(lateral epicondyle)
• Moving arm – aligned with metacarpal
of middle finger
Normal Range
• 35° ± 3.8° (American Academy of
Orthopaedic Surgeons)
• 30° (American Medical Association)
• 36.0° (mean) 3.8° (standard
deviation), (Boone and Azen)
METACARPOPHALANGEAL JOINT
FLEXION
Test Position
• Subject sitting with forearm resting on
table
• Wrist and interphalangeal joints
relaxed
• Forearm neutral
• Stabilize metacarpal to prevent motion
Goniometer Alignment
• Dorsal metacarpophalangeal joint
• Stationary arm - aligned with
metacarpal
Normal Range
• 86° (index), 91° (ring), 105° (little)
(American Academy of Orthopedic
Surgeons - active motion)
• 90° (American Medical Association)
METACARPOPHALANGEAL JOINT
EXTENSION
Test Position
• Subject sitting with forearm resting on
table
• Wrist and interphalangeal joints
relaxed
• Forearm neutral
• Stabilize metacarpal to prevent motion
Goniometer Alignment
• Dorsal metacarpophalangeal joint
• Stationary arm - aligned with
metacarpal
Normal Range
• 22° (index), 18° (long), 23° (ring), 19°
(little) (American Academy of
Orthopedic Surgeons - active motion)
• 20° (American Medical Association)
METACARPOPHALANGEAL JOINT
ABDUCTION
Test Position
• Subject sitting with forearm resting on
table
• Wrist neutral
• Forearm neutral
• Stabilize metacarpal to prevent motion
Goniometer Alignment
• Dorsal metacarpophalangeal joint
• Stationary arm - aligned with
metacarpal
• Moving arm – aligned with proximal
Normal Range
• 25° (Physiopedia)
METACARPOPHALANGEAL JOINT
ADDUCTION
Test Position
• Subject sitting with forearm resting on
table
• Wrist neutral
• Forearm neutral
• Stabilize metacarpal to prevent motion
Goniometer Alignment
• Dorsal metacarpophalangeal joint
• Stationary arm - aligned with
metacarpal
• Moving arm – aligned with proximal
Normal Range
• Either 0° or usually not measured.
INTERPHALANGEAL JOINT FLEXION
Test Position
• Subject sitting with forearm resting on
table
• Wrist, metacarpal, and non-tested
interphalangeal joints relaxed
• Forearm neutral
• Stabilize proximal bone to prevent
motion
Goniometer Alignment
• Dorsal proximal interphalangeal joint
• Stationary arm - aligned with proximal
phalange
• Moving arm – aligned with middle
phalange
Normal Range
American Academy of Orthopedic
Surgeons
• PIP fingers - 102° (index), 105° (long),
108° (ring), 106° (little) (active motion)
• DIP fingers - 72° (index), 71° (long),
63° (ring), 65° (little) (active motion)
• IP thumb - 73°
American Medical Association
• 100° (PIP finger), 70° (DIP fingers),
80° (IP thumb)
INTERPHALANGEAL JOINT
EXTENSION
Test Position
• Subject sitting with forearm resting on
table
• Wrist, metacarpal, and non-tested
interphalangeal joints relaxed
• Forearm neutral
• Stabilize proximal bone to prevent
motion
Goniometer Alignment
• Dorsal proximal interphalangeal joint
• Stationary arm - aligned with proximal
phalange
• Moving arm – aligned with middle
phalange
Normal Range
American Academy of Orthopedic
Surgeons
• PIP fingers - 7° (index), 7° (long), 6°
(ring), 9° (little) (active motion)
• DIP fingers - 8° (all finger DIPs)
• IP thumb - 5°
American Medical Association
• 0°
THUMB CARPOMETACARPAL JOINT
FLEXION
Test Position
• Subject sitting with forearm resting on
table
• Wrist neutral
• Stabilize carpals to prevent wrist motion
Goniometer Alignment
• Axis - carpometacarpal joint
• Stationary arm - aligned with radius
• Moving arm – aligned with metacarpal of
thumb
Normal Range
• 15° (American Academy of Orthopedic
Surgeons)
THUMB CARPOMETACARPAL JOINT
EXTENSION
Test Position
• Subject sitting with forearm resting on
table
• Wrist neutral
• Stabilize carpals to prevent wrist
motion
Goniometer Alignment
• Axis - carpometacarpal joint
• Stationary arm - aligned with radius
• Moving arm – aligned with metacarpal
of thumb
Normal Range
• 20° (American Academy of Orthopedic
Surgeons)
THUMB CARPOMETACARPAL JOINT
ABDUCTION
Test Position
• Subject sitting with forearm resting on
table
• Wrist neutral
• Forearm neutral
• Stabilize carpals to prevent wrist
motion
Goniometer Alignment
• Axis - radial styloid
• Stationary arm - aligned with
metacarpal of index finger
• Moving arm – aligned with metacarpal
of thumb
Normal Range
• 70° (American Academy of Orthopedic
Surgeons)
THUMB CARPOMETACARPAL JOINT
ADDUCTION
Test Position
• Subject sitting with forearm resting on
table
• Wrist neutral
• Forearm neutral
• Stabilize carpals to prevent wrist
motion
Goniometer Alignment
• Axis - radial styloid
• Stationary arm - aligned with
metacarpal of index finger
• Moving arm – aligned with metacarpal
of thumb
Normal Range
• Either 0° or usually not measured.
NOTE: Thumb adduction is the return to neutral from thumb
abduction. Thumb adduction is rarely measured, probably because it
THUMB CARPOMETACARPAL JOINT
OPPOSITION
Test Position
• Subject sitting with forearm
supinated and resting on table
• Wrist neutral
• Stabilize fifth metacarpal
Goniometer Alignment
• Goniometer cannot be used
• Use a ruler to measure distance
between tip of thumb and base of
fifth finger
Normal Range
• Able to touch tip of thumb to base
of fifth finger (American Academy
of Orthopedic Surgeons)
NOTE: The American Academy of Orthopedic Surgeons suggests
that opposition range is normal when the tip of thumb can touch the
base of fifth finger. When range is not adequate, a ruler can be used
to measure the distance between the tip of thumb and the base of the
NORMAL ROM VALUES (AAOS VALUES)
FINGERS: COMPOSITE FLEXION OF
THE MCP, PIP, AND DIP JOINTS
GRAVITY
GONIOMETER/INCLINOMETER
One arm has a weighted pointer that remains vertical under the
influence of gravity
BUBBLE INCLINOMETER
SOFTWARE/SMARTPHONE-BASED
GONIOMETER APPS
• Goniometer Records
• Goniometro Advance
• Goniometer Pro App (G-Pro)
• Goniometer Plus
• DrGoniometer
DIGITAL LASER GONIOMETER
THE E-LINK ROM GONIOMETRY
ELECTRO GONIOMETER
(Fariborz et al.,
2015)
ELECTROGONIOMETER
SIN GLE A XIS
ELEC TR OGON IOMETER
TW IN A XIS
ELEC TR OGON IOMETER
FIBER OPTIC GONIOMETRIC
GLOVE
(Sam Wise et al.,
1990)
CYBERGLOVE – VIRTUAL
GONIOMETRY
G. Drew Kessler et al.,
NON-CONTACT MEASUREMENT SYSTEM
(INTEL PERCEPTUAL TECHNOLOGY AND A SENZ3D CAMERA)
(Trieu Pham et al.,
2015)
3D MOTION ANALYSIS
(Gabriella Fischer et al.,
REFERENCES
• American Academy of Orthopedic Surgeons (Green WB, Heckman JD eds.): The Clinical Measurement
of Joint Motion. 6300 North River Road, Rosemont, IL 60018, 1994, ISBN 0-89203-090-9.
• American Medical Association: Guide to the Evaluation of Permanent Impairment. AMA, Chicago, 1988.
• Boone DC, Azen SP: Normal range of motion of joints in male subjects. Journal of Bone and Joint
Surgery 1979;61A:756-759.
• Magee DJ: Orthopedic Physical Assessment, 3rd ed. WB Saunders Co, Philadelphia, 1997, ISBN 0-
7216-6290-0.
• Norkin CC, White DJ: Measurement of Joint Motion: A Guide to Goniometry, 2nd ed. FA Davis Co,
Philadelphia, 1995, ISBN 0-8036-6579-2.
• Physiopedia