 Atomic Absorption Spectroscopy (AAS) is one of the
most popular and useful methods of elemental analysis
that measures radiations absorbed by neutral (unexcited)
atoms in the gaseous state.
 This technique was developed by Alan Walsh in 1956. It is
a powerful instrumental technique which is used for
quantitative estimation of trace metals in liquid samples.
 The technique which is used for quantitative estimation of
trace metals in liquid samples.
 The technique is very versatile as it has helped in
determining 60-70 trace metals whose concentration
ranged from trace to macromoles.
 It can be used in for the estimation of both aqueous and
non-aqueous solutions. Moreover, the technique is also
helpful in estimating a particular element even in the
presence of several other elements.
 This implies that it does not require separation of the test
element from the rest which saves a great deal of time also
helps to overcome the possibility of errors.
Characteristics of AAS
 It is a highly sensitive and error free technique.
 It is most suited method for the analysis of metals like Al, As,
Ba, B, Ca, Cu, Pb, Ni, K, Na, Se, Sn, Te, Zn etc.
 It is applicable to aqueous as well as non-aqueous solutions.
 It requires less time.
 It does not require sample preparation.
 It requires very little amount of sample.
 It is independent of the molecular form of the metal in the
sample. For example, the total sodium content of water sample
can be determined irrespective of the molecular form (halide,
sulphate, hydride) in which the sodium exists.
Advantages of AAS over Flame Photometry
 It is less liable to spectral interferences when
compared to flame photometry.
 It is a more reliable method.
 It is relatively more sensitive than flame photometry
in the detection of Ag, As, Au, B, Bi, Cd, Co and Fe.
 Since large no. of atoms produce atomic absorption
signal and as only thermal energy is utilized to
atomize the sample, AAS is independent of flame
temperature.
 It is a more specific technique because the atoms of a
particular element can absorb the radiation of their
own characteristic wavelength.
 It obeys Beer’s law over a wide concentration range
which is not true in case of flame photometry.
 A wide range of elements can be quantitatively
determined by this technique.
 It is a more specific process because wavelength of
absorbed radiation is characteristic of elements.
 It requires only small amounts of sample.
Disadvantages:
 A separate lamp is required for the determination of each
element. This difficulty can be overcome either by using a
continuous source of Electro Magnetic Radiation (EMR) with a
high resolution monochromator or by using a line source of
EMR with the help of flame.
 Elements like aluminium, molybdenum, vanadium, titanium
and silicon etc., form red metallic oxides in the flame and
hence cannot be detected using AAS, except under modified
conditions.
 When aqueous solutions are used, the predominant anion
affects the absorption signal to a high degree.
 This technique cannot be used for the detection of non metals
due to the strong absorption of light by flame gases.
Principle
 The basic principle of atomic absorption spectroscopy (AAS) is
the absorption of energy by the ground state atoms in the
gaseous form.
 When a solution containing a specific metal is introduced into a
flame, it results in the formation of vapourized forms of the
metallic species.
 Only some of the metal atoms from the ground state undergo
transition to such an energy level that they are able to emit the
radiation which is characteristic to the metal.
However, most of the metal atoms remain in the non
emitting ground state. These non-emitting ground
state atoms absorb that light radiation which has the
wavelength specific to them.
Therefore, when light of this specific wavelength is
passed through a flame which contains the atoms of
the metallic species, then some of the light will be
absorbed and the absorption is directly proportional to
the density of the atoms of metallic species present in
the flame.
Hence, in AAS the concentration of the test metal is
directly determined from the amount of light
absorbed.
The total amount of light absorbed can be mathematically expressed
as,
The total amount of light absorbed(at ύ)= πe2/mc X Nf 1
Where,
ύ = Frequency of the light
e = Charge on electron
m= Mass of electron
c = Speed of light
N = Total no. Of atoms that can absorb at ύ
f = Oscillator strength (Ability of an atom to absorb at ύ)
Since, π, e, m and c are constants, the above equation gets simplified
as,
Total amount of light absorbed = KNf 2
From equation 2, it is clear that the total amount of light absorbed is
independent of wavelength of absorption and temperature.
Instrumentation
Instrumentation
Components of Atomic Absorption Spectrophotometer
Radiation Source
 AAS is concerned with the transition of free atoms of the test
metal from their ground state to the excited state.
 The transition from the ground state to the first excited state is
known as the first resonance line.
 It is the line with strongest absorptivity.
 The absorptivity of a particular element decreases as the energy
difference between the ground state and the excited state
increases.
 When high sensitivity is required, the first resonance line of the
test element is used.
 The wavelength of first resonance line for all metals and
metalloids is longer than 200nm and for non-metals it lies in
vacuum UV region below 185nm and hence cannot be
measured by conventional spectrophotometers.
 Atomic absorption spectrophotometers are the widely used for
analysis of metals and metalloids and the non- metals can be
detected by modifying the optical system of the
spectrophotometer.
 However, such modifications make the instrument expensive
and hence these are not widely used.
For effective quantitative estimation by AAS, the radiation
source should emit the resonance spectral lines of the test
metal.
The bandwidth of the incident light from the radiation
source should be narrower than the absorption line of the
sample.
This implies that the bandwidth of the radiation source
should be less than 0.001nm.
 Moreover, there should not arise any extraneous line
emitting within the band pass of the monochromator.
 The problem of narrow resonance spectral lines can be
overcome by the use of hallow cathode lamp.
Hollow cathode lamp:
 It is the most commonly used radiation source for atomic
measurements.
It consists of an evacuated glass tube with a tungsten
anode and a cylindrical cathode.
 The hollow cathode is made up of the element to be
analyzed or its alloy and is enclosed by porcelain or glass
to avoid diffusion of the atoms produced.
The glass tube is filled with an inert gas.
Application of a potential of about 100-200 volts across
the electrodes causes the ionization of inert gas.
 The charged gas then migrates at high velocity towards
the cathode and collides with the metal atoms from its
surface.
The metal atoms absorb energy and get excited to higher
energy levels.
The excited atoms then transit to the lower energy states
and emit radiations characteristic of metal used in cathode.
 Therefore, the emission spectrum given by the hollow
cathode lamp is actually a sharp line spectrum of the
cathode metal and inert gas.
 The spectral lines produced by the hollow cathode
lamp are narrower than the absorption line of the
source.
 For example, a zinc cathode emits a zinc spectrum
and a copper cathode emits a copper spectrum.
Disadvantages
 Different hollow cathode lamps are required for the
determination of different elements.
 Attempts to increase the intensity of emitted radiation
results in increased sputtering, resonance broadening
and self-reversal phenomenon (absorption of emitted
spectrum by similar atoms in lamp)
 It restricts the analysis of multiple elements.
Electrodeless Discharge lamps
 Stable hollow cathode lamps are difficult to construct
from volatile elements like selenium, germanium and
arsenic.
 Hence for the determination of such volatile elements,
electrodeless discharge lamps have been developed.
 Each lamp consists of a sealed quartz tube filled with an
inert gas (argon) and in which the test metal is placed.
The lamp is placed in a microwave radiation field so that
the sample element vaporizes and transits to higher energy
levels.
As the excited atoms return to lower or ground state,
radiation that is characteristic of the metal element is
emitted.
 Intensity of the emitted radiation is 10 times more than
that obtained using cathode lamp.
Atomizers
 Flame atomizers like total consumption burner and
premix burners are used for converting the liquid
sample into gaseous form and also for converting the
sample molecules into atomic vapours.
Total consumption burner/Pneumatic Nebulizer
 This burner uses hydrogen (fuel) along with oxygen (oxidant)
to produce flame.
 In it, the liquid sample, fuel and oxidant are introduced through
different inlets.
 When the liquid sample reaches the base of the flame, it gets
converted into fine droplets and then gets burnt or evaporated.
 The residue left is further reduced into atoms which are then
excited.
Advantages
 The total amount of sample entering the inlet aspirates
completely, irrespective of its droplet size.
 Simple in design.
Disadvantages
 Burner produces a lot of noise.
 Flame is not homogeneous which results in variation
in the intensity of the flame.
 Less efficient.
 Rate of introduction of sample is dependent on its
viscosity.
Premix Burner/Laminar Flow Burner
 Premix burner is also called Laminar flow burner because the
gases move in laminar (parallel) or non-turbulent fashion.
 In this, the liquid sample is initially aspirated with air and then
thoroughly mixed with fuel and oxidant.
 The mixture then reaches the opening of the burner and enters
into the flame.
 Only about 5% of the sample in the form of fine droplets
reaches the flame whereas large droplets i.e., 95% of the
sample drains out from the side of the spray chamber (burner).
 This small portion of the sample is easily decomposed in the
flame and helps in effective atomization of the sample.
Fig: Premix Burner
Advantages:
 Flame produced is stable.
 It is a quite process i.e., the burner does not produce noise.
 Non-turbulent flame is produced.
 Effective atomization of the sample takes place.
 Clogging of the sample solution does not occur.
Disadvantages
 Chances of explosion in the mixing chamber of the burner.
 Sensitivity is low because most of the sample is wasted.
Fuel and oxidants
 Air-acetylene or nitrous oxide-acetylene are the most
commonly used combination of fuels and oxidants.
 Temperature of the fuel affects the absorption spectra in a
complex manner.
 Lower temperatures prove to be inefficient in elevating
atoms to an excited state while higher temperatures tend to
cause ionization of the atoms.
 Thus, the desired no. Of may not be achieved in both the
cases.
 Hence, it is necessary to choose an ideal combination of
fuel and oxidant that gives the required temperature.
Mixture of Fuel and
Oxidant
Temperature (0C)
Hydrogen + Oxygen 2800
Hydrogen + Air 2100
Acetylene + Oxygen 3000
Acetylene + Air 2200
Propane + Oxygen 2800
Propane + Air 1900
Butane + Oxygen 2800
Butane + Air 1900
Cyanogen gas + Oxygen 4500
Table: Fuel and Oxidant Mixtures
4.Monochromator
5. Detector
Photomultiplier tube is the most suitable detector for atomic
absorption spectroscopy.
6.Amplifier
 Lock-in amplifiers are suitable for atomic absorption
spectroscopy.
 An AC- amplifier modulates the signal from reference
source at the same frequency and directs each of them to
the synchronous detector (via a phase shifter) where in the
two signals combine and generate sum and difference
frequencies.
 The signal then enters into the low pass RC filter that
eliminates the unwanted signals and directs the signal of
interest to DC amplifier from where it enters into the read
out device.
Block diagram of Lock-in amplifier
7.Read out device
 Chart recorders are mostly used as readout devices in
atomic absorption spectroscopy. Digital readout devices
are also available.
Working of Atomic Absorption Spectrophotometer
 Initially the detector-amplifier unit is adjusted to read
100% transmittance or zero absorbance.
 A blank solution is then sprayed into the flame and
simultaneously the uninterrupted beam of light from the
lamp is allowed to fall onto the Photomultiplier tube.
 When the solution containing test material is introduced
into the flame and illuminated, a portion of the radiation is
absorbed by the excited atoms.
 This causes a decrease in intensity of light falling onto the
photomultiplier tube which produces a deflection and
yields the value of absorption.
 A calibration curve is drawn by preparing standard
solutions of the element to be determined and determining
their absorbance.
 The concentration of test metal in the sample can be
measured by extrapolation of the calibration curve.
Single Beam Atomic Absorption Spectrophotometer
 In single-beam atomic absorption spectrophotometer, chopper
is placed between the hollow cathode lamp and flame.
 When a blank solution is sprayed into the flame and radiation
from the lamp is allowed to fall onto the detector, the detector-
amplifier unit is adjusted to give zero absorbance or full
transmittance.
 When the sample solution is sprayed into the flame, there
occurs a decrease in the intensity of light and the output of
detector-amplifier unit is recorded.
 One of the demerit of single beam equipment is its low
stability.
Double beam Atomic Absorption Spectrophotometer
 The Double beam Atomic Absorption Spectrophotometer
overcomes the demerit of single beam equipment.
 In this, beam of light from the source is split into two, one
portion is allowed to pass through the flame while the
other portion is not allowed.
 The two beams of light are recombined and made to pass
through the monochromator and finally to the read out
device.
Applications:
Qualitative Analysis
 AAS is not used for Qualitative analysis since separate lamps
are required for the determination of each element.
Quantitative Analysis
 Quantitative Analysis of various samples by AAS is achieved
by Calibration curve method. In this method, a series of
standard solutions of known concentration of the element to
be analyzed is prepared.
 The concentrations of standard solution are plotted on X-axis
and their respective absorbance on Y-axis. The absorbance of
sample solution is measure, plotted on graph and then
extrapolated on X-axis to get the concentration of the
element being analyzed.
 Concentration up to 0.001ppm can be determined using
AAS.
 It is the most suitable method for the estimation of sodium,
potassium, calcium, magnesium and lithium in serum and
plasma and lead, iron and zinc in whole blood. Determination
of these elements helps in diagnosing pathological states.
 It is an important tool for the estimation of elements like
copper, nickel and zinc in food products.
 It is an ideal method for the analysis of alloys, minerals, ores
etc., and hence is widely used in metallurgy and inorganic
analysis.
Helps to estimate pollutants like cadmium, chromium,
copper, lead, mercury, nickel and zinc in water, soil
and environmental samples.
 Estimation of Zinc in zinc insulin injection, oils, creams,
calamine lotion, calcium in calcium preparations, lead in
CaCo3 and mercury in thiomersal solution is possible with
AAS.
 AAs is also helpful in estimating the levels of Na+, K+,
Ca2+ in saline and Ringer’s solution.
 It has been used for the assay of cisplatin (for Ag),
intraperitoneal dialysis fluid (for Ca, Mg), activated
charcoal (for Zn) etc.
In cement industry, AAS is used to determine the quality of
cement as it helps in estimating sodium, potassium,
magnesium and calcium content of cement.
The boiler ash is analyzed for sodium, potassium,
calcium, magnesium and iron by AAS.