BUSHRA IQBAL(11740)
HISTORY
        The beautiful phenomenon of “RAINBOW”
            was the first dispersed spectrum.

In 1665 NEWTON took the first and the most important step
        towards the development of spectroscopy.




In 1859 G.R KIRCHOFF and R BUNSEN emerged as the FATHER
               OF MODERN SPECTROSCOPY.
SPECTROSCOPY
  The study of structure of atom or molecules.
    Used in analysis of wide range of sample
Can classify in atomic or molecular spectroscopy.
ATOMIC SPETROCOPY




MOLECULAR SPECTROSCOPY
ABSORPTION SPECTROSCOPY
• It gives the measurement of absorbed energy by the
  atom or molecule after which it excite and gives the
  absorbed energy in the form of absorption spectrum
ATOMIC ABSORPTION SPECTROSCOPY
• Atomic absorption spectroscopy is a technique for
  determining the concentration of a particular metal
  element in a sample. Atomic absorption
  spectroscopy can be used to analyze the
  concentration of over 62 different metals in a
  solution.
TYPES OF ATOMIC SPECTRA
1) ATOMIC EMISSION SPECTRA:
2) ATOMIC ABSORPTION SPECTRA:
3) ATOMIC FLUORESCENCE SPECTRA
COMPARISON

ATOMIC EMISSION                       ATOMIC ABSORPTION
SPECTROSCOPY                          SPECTROSCPY
• Examines the wavelengths of         • Measures the loss of
  photons emitted by atoms or           electromagnetic energy after it
  molecules during their transition     illuminates the sample under
  from an excited state to a lower      study.
  energy state.                       • The energy in certain amount
• Each element emits a                  is absorbed during transition
  characteristic set of discrete        to the higher level.
  wavelengths.                        • The amount of energy
• By observing these wavelengths        absorbed gives estimate of the
  the elemental composition of          concentration of the analyte in
  the sample can be determined.         the sample.
PRESENTORS:

RIMSHA ISMAIL (11780)
RUTABA MURTAZA (11782)
THEORY OF ATOMIC ABSORPTION
           SPECTROSCOPY




•   Absorption therefore is carried out by unexcited atoms.
CONCEPT OF ATOMIC ABSORPTION
                  SPECTROSCOPY
• In 1955, A. WALSH and another one of C T J ALKAMADE and J M W
  MILATZ.
PRINCIPLE OF ATOMIC ABSORPTION SPECTROSCOPY
                                            EVAPORATION                VAPORIZATION
 SOURCE                    SOLUTION                             MIST




           DISSOCIATION                            GAS                      SOLID

                                              hv
                 THERMAL                                     FLAME
                EXCITATION                                  EMISSION

                                      hv2          RE-EMISSION
       ABSORPTION OF
       RADIANTENERG                                (FLUORESCENCE)
       Y


RECORDER                               DETCETOR           MONOCHROM
               AMPLIFIER                                  ATER
• The total amount of light absorbed may be given mathematically by the
  following expressions:

               Total number of light absorbed = πe2/mc Nf
Where,
  e= is the charge on the electron of mass
  m= mass of electron
  c= is the speed of light
  N= is the total number of atoms that can absorb at frequency in the light
 path
  v= frequency
  f= is the oscillator strength or ability of each atom to absorb at frequency
 π= is constant

The above equation can be written as:

             Total amount of light absorbed= Constant x N x f
INSTRUMENTATION OF ATOMIC
 ABSORPTION SPECTROSCOPY
1. RADIATION SOURCE

The main sources used for atomic absorption are:
I. Hollow Cathode Lamp (HCL)
II. Electrode less Discharge Lamp (EDL)
I.   HOLLOW CATHODE LAMP
II. ELECTRODELESS DISCHARGE
              LAMP
SOURCE MODULATION
2. ATOMIZERS




           There are two types of atomizers which
             are used in Atomic Absorption
             Spectroscopy.
           • Flame Atomizers
           • Electrothermal Atomizers
FLAME ATOMIZER:
• Flame is used to atomize the sample
• Sample when heated is broken into its atoms




PRINCIPLE:
• High temperature of flame causes excitation
• Electrons of the atomized sample are promoted to higher
  orbitals, by absorbing certain amount of energy


QUANTITATIVE ANALYSIS:
• The amount of energy absorbed is specific for a particular element
  (for electronic transition).
• Amount of absorbed radiation is a quantitative measure for the
  concentration of the element to be analyzed
SAMPLE             NEBULIZER
                      ASSEMBLY
         FLAME
     (ATOMIZATION
        OCCURS)


                      CONVERSION
                     INTO FINE MIST
  AEROSOL
                        & SMALL
 MIXES WITH
                      DROPLETS OF
COMBUSTION
                       SOLUTION
   GASES
               ASPIRATED
              INTO SPRAY
               CHAMBER
                (MIXING
               CHAMBER)
Sample to be       Nebulizer:
 nebulized is
                  sample+fuel+
    taken
 (aspirated)       oxidant
via a capillary   fine mist or
     tube            aerosol
Natural Gas        Air        1700-1900     39-43

Natural Gas     Oxygen        2700-2800    370-390

Hydrogen           Air        2000-2100    300-440

Hydrogen        Oxygen        2550-2700   900-1400

Acetylene          Air        2100-2400    158-266

Acetylene       Oxygen        3050-3150   1100-2480

Acetylene     Nitrous Oxide   2600-2800     285
- Wavelength selectors
  - Produces
  monochromatic light

Consists of:
1) Entrance slit
2) Diffraction grating
3) Exit slit


 Diffraction gratings are
 mostly used rather than
 prisms
Electro thermal atomization (ETA) (also known
as Graphite furnace atomization)



Used for improvement and betterment in the limit-
of-detection and sensitivity for atomic absorption
measurements.
GRAPHITE FURNACE ATOMIZERS:
Graphite tube

Enclosed water
cooled housing
  Transparent
    windows
Inert purge gas
     control
Electrical contact
1.Drying


Atomization    2. Ashing or
 of sample      pyrolysis


              3.Atomization
WORKING OF PHOTOMULTIPLIER TUBE
V ideo0003.3gp
The signal could be displayed for readout in the readout devices.




            Readout
            devices:
• Digital voltmeter
• Simple galvanometer
• Potentiometer
• Computer
• Atomic Absorption spectrophotometric measurements are
  done extensively by using;
 Single-Beam AA Spectrophotometer
 Double-Beam AA Spectrophotometer
• The first AAS was presented by Walsh and co-workers in
  Melbourne in 1954, was a double beam atomic absorption
  spectrophotometer.
• Walsh worked with Perkin-Elmer, the first AAS instrument
  developed by that company was MODEL 303.
Single Beam AA Spectrophotometer:

   Single beam measurements are depended upon the varying
intensity of a single beam of light having a single optical path.
That is why called as single beam AA spectrophotometer.
Double Beam AA Spectrophotometer:

• It split the light from the source into a ‘sample beam’
  (focused to the sample cell) and a ‘reference beam’ (focused
  around the sample cell).
• Such an instrument is called as double beam AA
  spectrophotometer, as measurements are made on varying
  intensity of double beams of light in dual optical path.
Single Beam Instruments                   Double Beam Instruments
 Simple, Less expensive, less complexity     Complex, more expensive, not easily made

 Low automation, more efficiency of light    High speed automation, less light efficiency

 More time consuming                         Less time consuming

 Low stability                               Increased stability

 Depend upon single beam intensity           Depend upon ratio between both beams

 More chances of fluctuations                Lesser fluctuations in readings

Sample and Reference placed separately       Sample and Reference are kept at same
                                             time
High light throughput, more resolution       Less light throughput, decreased resolution

More warm-up time is required               Less or reduced warm-up time required
SAMPLE PREPARATION
• The preparation of the sample solution for a solid
material is most time consuming step of process
 of analysis in an atomic absorption spectroscopy.
It involves following steps;


        Weighing of sample


      Dissolution in appropriate solvent or
     digestion using different techniques


                Dilution of sample if necessary
• A sample employed for atomic absorption spectroscopy in the
   laboratory is placed into one of the following categories:
• Considerations are to be given to
some of the general principles involved
in the various sample preparations.
• A little preparation is required with this sort of sample.
• These include samples in raw and treated water, sea
  waters, biological fluids, beer, plating
  solutions, effluents, wines etc.
• It include petroleum products, many of which can be directly
  be aspired. Examples of such solvents used are m-
  heptene, aliphatic ketones, (e.g. methyl iso butyl ketone)
  aliphatic esters, alcohols and
  xylene, cyclohexanes, isopropanol etc. are frequently
  employed.
• Note: When samples are analyzed in organic solvents
  some adjustments such as ‘BURNER
  CONTROLS’,               proper ventilation or other
  appropriate settings              must be used.
•  These include solid samples of fertilizers, ceramics, alloys or
  rocks.
• These are solvated by using appropriate aqueous or acid
  medium depending on solubility. Such as hot
  water, concentrated acids, acidic mixtures, dilute acids etc.
• Other techniques can also be employed such as fusion
  prolonged acid digestion, wet ashing etc to yield sample
  solutions.
• These includes typically material of foods, leaves, tissue, biological
  solids, polymers, plants, feedstuff etc.
• Before solublization of such samples, there is a requirement of
  destruction via wet digestion or ashing in a furnace (muffle).
•   Atomic absorption analytical works
or procedures can be employed in
analyzing gases indirectly as
liquid sample.
PREPARATION:

    Separate metals         Using Millipore         Wash or
       from gas                filter disc          dissolve




                           Analyze using
                                              Using nitric acid
                            standards
Techniques Used For Certain Sample
           Preparations:
• As discussed in previous topics, some solid sample requires
  special techniques for dissolution. Such sample preparation
  requires time and proper handling.
Wet Ashing Or Wet Digestion:

                         Undissolved in
Inorganic samples                               Treated with acids
                        aqueous solvents




                                               Clean liquid with no
Such as Perchloric        Digestion of
                                                  single element
 acid, HCl, HNO3,      complexes, silicates
                                                  being removed
Dry Ashing:
           Heated in   Mix left   Remove inorganic
Weighed
            furnace    over in    particles like lead
sample                               or mercury
            (muffle)    acid



Microwave Dissolution: (type of digestion)
Extraction and Concentration process:
• Such an operation is done if sample contain species
  interfering in absorption or the concentration of sample
  desired is in low concentration to show absorption readings.
  Such a process involves,
•       The viscosity adjustment can be
 done with suitable solvents which can be
that in a way solvent should;
o Dissolve or mix completely with the sample
o Well burnt but in a controlled manner
o Be in pure state such that possessing no species having
   molecular absorption in region
o Not yield harmful by-products
o Not be expensive
PRESENTOR:
  Shumaila Eliyas
“Increase or decrease in the size of the signal
  obtained from the analyte as a result of the
  presence of some other known or unknown
  component in the sample”
 Types of interference:
 Spectral interference
Chemical interference
Spectral interference
 This may be caused by direct overlap of the
 analytical line with the absorption line of the
 matrix element.
HOW TO OVERCOME ?
By choosing an alternate analytical
 wavelength
By removing the interfering element from the
 sample.
Examples of spectral interferences
Chemical interference

Formation of compound
    of low volatility

  Decrease in calcium
absorbance is observed
    with increasing
concentration of sulfate
     or phosphate
 By increasing flame temperature
Use of releasing agents (La 3+ )
 Cations react with the interferent releasing
 the analyte
Use of protective agents:
 They form stable but volatile compounds with
 analyte.
Ionization interference
Ionization of ground state gaseous atom with in a flame
  will reduce extent of absorption in AAS.
                 M ↔ M+ + e

HOW TO MINIMIZE:
 Low temperature of the flame

 Addition of an excess of ionization suppressant e.g.
  the alkali metals (K, Na, Rb, and Cs)
Influence of potassium on the
     ionization of barium
ATOMIC ABSORPTION SPECTROSCOPY
 Pharmaceutical

     Biological

     Biochemical


For detection of purity and
consistency of these trace metals

Also for quantitative determination
of metals mainly in solid sample as
mineral, ores and alloys
• Magnesium in cast iron

• Silver, Zinc, Copper and Lead in
Cadmium metal




• Method of multiple standard addition
• A plot of absorbance against the amount
of standard can be used to determine the
amount of copper in a sample.
 Determination of trace metal in a silicon foam cavity wound dressing


 Zinc in Zinc insulin suspension and tetracosactrin Zinc
injection

 Copper and Iron in ascorbic acid

 Aluminum in albumin solution and Ca, Mg,
  Mercury

 Zinc in water used for diluting haemodialysis solution
•For the analysis of pharmaceutically or therapeutically
essential component of formulation, such as Zinc in Zinc-
insulin, minerals in multivitamin-mineral preparation and
Ca, Mg, Al in antacids.

•To establish concentration limits where the metal is
regarded as an impurity.
•Mining industries

•Petroleum industries

•Determination of metallic elements in
food industry like Copper, Zinc and Nickel
in vegetable oil and copper in beer.